Doc -030 Hatch comfort C1 User manual
Revised on 15/06/2021
10.1.2
Disinfection with alcohol
CAUTION
Use alcohol for disinfection of joints, axis and switches.
▶
Other types of cleaning solution may cause discoloration or damage to the
chair materials.
Do not let alcohol smear into the dental chair unit.
A special attention is required on joints, axis and switches.
Also do not use excessive amount of alcohol.
▶
Otherwise, short circuit, malfunction, fatigue failure and damage on material
may occur.
※
As there are different requirement among models, refer to the manual of each instrument.
▶
Disinfection of the operating light reflector
CAUTION
The operating light is hot after use. Cleaning and disinfection should be
performed after it cools down completely.
1)
Loosen the cover fixation screw and
take off the cover.
2)
Disassemble the lamp cap by turning it
counterclockwise.
3)
With soft fabric soaked in alcohol,
wipe off front and back gently.
4)
Assemble parts in the reverse order.
Light
cover
Fixation hook
for cover
Lamp cap
Reflector
Lamp
socket
Cover fixation screw
Lamp
CAUTION
Do not touch the halogen lamp in any case.
Wipe the surface gently so coating on the reflector surface may not peel off.
NOTE
Other parts of the operating light should be wiped off with gauze or soft fabric
soaked in alcohol. Excessive alcohol in gauze should be squeezed out.