3-WAY syringe
Pick up syringes from both the dentist and assistant holder and press levers to eject
water into the spittoon bowl.
▶
Air turbine and micromotor
1)
Pick up a handpiece from the holder.
2)
Disconnect the handpieces from tubing.
3)
Point its tubing to the spittoon bowl and press the foot controller pedal.
Keep this action for 2 minutes at least.
4)
Connect the handpiece to tubing. Put back the handpiece to the holder.
5)
Perform the same procedure with other handpieces.
▶
Scaler
CAUTION
A scaler should be operated with a tip connected all the time.
▶
Otherwise, it may overload the circuit or other part causing breakdown or
malfunction.
1)
Pick up the scaler from the holder.
2)
Point the scaler tip to the spittoon bowl and press the foot controller pedal.
Keep this action for 3 minutes at least.
If the water spray does not work, press WATER SPRAY switch in the control panel.
3)
Put back the scaler to the holder.
10.1.8
Cleaning other parts and surface
WARNING
Do not let water, chemical solution and alcohol remain on the surface and inside
of the dental chair unit. If they smear into the dental chair unit, turn off the
main switch and dry the dental chair unit completely.
▶
Otherwise, water may ingress inside of the dental chair unit and cause
electrical short circuit and discoloration.
Use only neutral detergents (alcohol) or cleaning detergents recommended by
Hatch Group Inc..
▶
If wiped with fabric soaked in benzene, thinner or other cleaning chemicals
other than recommended, parts may get damaged by discoloration.