Doc -030 Hatch comfort C1 User manual
Revised on 15/06/2021
Adjustment cap
Illumination
range
X
Y
CAUTION
When the operating light is in use, all parts, except handle, may be hot. It may
cause burns if touched.
8.10.3
Focus adjustment
CAUTION
The range of the adjustment cap movement is limited.
If the cap movement gets stiff, stop turning it.
If turned forcibly, it may be damaged and dismantled.
▶ The operating light is configured at the
standard distance of 700mm from the
patient mouth. However, the cap can be adjusted quickly and easily between the
focus range of 600mm~1,000mm distance.
Adjustment knob for filament
700mm
8.10.4
Adjustment of light intensity in 2 steps (standard)
▶
Press the
switch on the dentist control panel
to turn on the operating light. Press it again to turn
off the operating light.
▶ Adjust the light intensity by turning a toggle switch
in the unit body.
Light intensity switch
▶
Adjust the direction of filament in lamp with
‘Filament
direction
adjustment
knob’
by loosening the knob a bit and turning it along with the guide
rail to rotate the lamp.
After controlling the direction of lamp, be sure to tighten the knob.
Depending on the degree of rotation, the illumination range and intensity change.