HARDY MANUFACTURING CO., INC (MODEL LC300, LC500 & LC2500)
PAGE 52
8-6
Automatic Water Fill Replacement Instructions (continued)
Step #
Operation Description
Visual Aids/Comments
6
Note: The automatic water fill valve has an arrow
stamped in the housing showing the direction of flow.
The valve has to be installed with the arrow in the
correct direction for the valve to operate properly.
7
Apply pipe thread sealant to the threads of the fitting
removed in step 5. Install the fittings and the new valve
assembly on the heater as shown. Care should be
taken to make sure no excess pipe sealant gets into the
screen during assembly. This wil result in reduced water
flow. The brass elbow should point slightly forward.
8
Using a pipe wrench or other suitable tool turn the tee
approx. 1/4 turn counter-clockwise until the assembly is
parallel with the rear of the furnace. Reinstall the flex
line fitting and tighten hand tight plus approx. 1/2 turn.
Do not over tighten the flex line to prevent damage to
the inner bushing.
9
Slide solenoid over valve stem and press firmly into
place. Make sure spring retainer clip on solenoid
latches into groove in top of valve stem.
10
If replacing the solenoid strip wires approx. 1/2" and
connect using wire nut suitable for 2 #16 AWG wires
(orange) as shown. These wires are not polarity
sensitive.
Direction of Flow
!
Hazardous Voltage! Disconnect Power to
Heater prior to performing this
maintenance operation.