31
7.
Put one cuvette from the
Cuvettes and 1000 µL
Pipet Tip Set into the black
cuvette holder.
8.
Put a blue pipette tip on
the blue pipet.
9.
Push in the operating
button on the pipet to the
stop.
10.
Put the tip in the
sample water 1 cm below
the surface. Slowly release
the operating button to pull
in the sample.
11.
Put the tip into the
cuvette and dispense the
liquid by gently pushing in
the operating button.
12.
Remove the tip from
the pipet and put in the
waste bag. Put the pipet in
the storage case.
13.
Put a yellow pipet tip on
the yellow pipet.
Use a new pipet tip for
each reagent.
14.
Do steps 9 to 12 to put
100 µL of CT Reagents 1, 2
and 3 into the cuvette.
15.
Open the luminometer
lid. Remove the cuvette
from the cuvette holder.
Gently tap the cuvette twice
to mix the solution. Put the
cuvette in the luminometer
cell.