102
1.
Open the luminometer
lid and remove any sample
that is in the cell. Close the
lid.
2.
Push
PROCEED
to
show the test status.
When the cell tests are
done, push
PROCEED
again.
3.
If testing D 2 values and
higher, set the 0.2 - 1.0 mL
pipette to 0.5 mL.
If testing D 1 values and
higher, set the 0.2 - 1.0 mL
pipette to 0.8 mL.
4.
Open the luminometer
lid.
5.
Put the test tube in
position B1 into the black
test tube holder in the
luminometer cell.
6.
Close the luminometer
lid. Push
MEASURE
.