67
12.
Open the lid of the
luminometer.
Remove the test tube from
the luminometer and put it
back in the LUMIStherm.
13.
Do steps 22 to 24 again
to measure the three other
test tubes in the correct
order (2, 3, and then 4).
The Inhibit% and rel. units
are shown on the screen.
Record the Inhibit% and
rel. units values for each
sample dilution on the
Screening Luminescence
Results Sheet.
14.
Use the color chart on
the Screening
Luminescence Results
Sheet to identify which
sample dilutions are toxic
(red) and which are
non-toxic (green).
Note:
The more of your results
in the red zone, the stronger
are the inhibitory affects of the
sample, the more critical is the
sample.
15.
Put the solution in the
test tubes into the waste
bottle.
16.
Put the test tubes into
the waste bag.