19
The Set-up Menu is shown.
5.
Select Set Measurement Range and push
ENTER
.
The current range is shown.
6.
Push
YES
to confirm.
7.
To change the range, push
CHANGE
8.
Push
STORE
to save the change.
The Set-up Menu is shown.
3.2.3.2 Luminescent Bacteria Test
If the measurement range is set to the 0–2000 light units and the measuring value
is marked with an * (e.g., 2010*) or the lumiometer shows Detector Overload, the
measurement is above the set measurement range. If this occurs, dilute the
bacterial stock suspension with Diluent and do the reading again.
To show or change the measurement range for the Luminescent Bacteria Test:
1.
Push
ON
(green button) to apply power to the instrument.
The instrument does built-in tests that make sure that the electronics and
software are operating correctly.
2.
When the built-in tests are done, push
PROCEED
.
The Main Menu is shown.
3.
Select LUMINESCENT BACTERIA TEST
and push
ENTER
.
The LBT Main Menu is shown.
4.
Select Set-Up and push
ENTER
.
The LBT Set-Up Menu is shown.
5.
Select Set Measurement Range and push
ENTER
.
The current range is shown.
6.
Push
YES
to confirm.
7.
To change the range, push
CHANGE
8.
Push
STORE
to save the change.
The Set-up Menu is shown.
Содержание Eclox
Страница 1: ...2886888 Eclox Rapid Response Test Kit User Manual 01 2018 Edition 5 ...
Страница 2: ...Visit us at www hach com ...
Страница 6: ...Table of Contents Table of Contents Page 6 EcloxTOC fm ...
Страница 14: ...14 ...
Страница 24: ...24 ...
Страница 46: ...46 ...
Страница 52: ...52 ...
Страница 110: ...110 ...
Страница 113: ...113 ...
Страница 114: ...Visit us at www hach com ...
Страница 120: ...120 ...
Страница 123: ...123 ...