This procedure is temperature sensitive.
1.
Remove the
luminescent bacteria test
reagent from the freezer.
Remove the reconstitution
solution from refrigerator.
2.
Remove the cap from
the reconstitution solution
bottle.
3.
Remove the foil seal and
rubber stopper from the
reagent bottle.
4.
Set the 1.0-5.0 mL
pipette to 1.0 mL.
5.
Put the end of the pipette
into a clean pipette tip.
6.
Put the tip of the pipette
into the reconstitution
solution and slowly pull in
1.0 mL.
7.
Put the tip of the pipette
into the luminescent
bacteria reagent bottle.
Quickly dispense the
solution into the reagent.
8.
Put the rubber stopper in
the reagent bottle. Swirl the
reagent bottle to mix.
9.
Cool the sample for at
least 15 minutes in a
refrigerator.