66
Luminometer is shown, select LIMIT Measure and Send to PC and push
ENTER
. The luminometer must be connected to a computer (refer to
Connect
the luminometer to a computer on page 23
).
•
To measure the luminescence and print the results on a printer, select LIMIT
Measure and Send to Printer and push
ENTER
. The luminometer must be
connected to a printer (refer to
Connect the luminometer to a printer on
page 21
).
6.
Open the luminometer
lid and make sure a sample
is not in the cell. Close the
lid.
7.
Push
PROCEED
to
show the test status. When
the cell tests are done,
push
PROCEED
again.
8.
Wait until the timer clock
completes 15 minutes.
9.
Open the luminometer
lid.
10.
Put the test tube in
position 1 (non-toxic blank)
into the black test tube
holder in the luminometer
cell.
11.
Close the luminometer
lid. Push
MEASURE
.
When the measurement is
complete (approximately
15 seconds), the
luminometer shows the
relative light units
measured.