11.12 Measure the light intensity of the test suspension
after the sample dilutions are added
Measure the light intensity of the test suspension (luminescent bacteria) after the
sample dilutions are added according to ISO 11348 using this procedure.
The Luminescent Bacteria Toxicity Test is a biological test method and the result is
therefore strongly temperature-dependent. ISO 11348 states that the test must be
done under temperature controlled conditions at 15 °C using a thermostat
(LUMIStherm, LTV053).
If the Luminescent Bacteria Toxicity Test is done at ambient temperature, record the
temperature. The results of tests done at different temperatures cannot be
compared directly.
To measure the light intensity of the test suspension after the sample dilutions are
added:
1.
Wait until the contact
time is completed.
2.
Open the luminometer
lid.
3.
Put the test tube in
position B1 into the black
test tube holder in the
luminometer cell.