Troubleshooting.fm
Page 115
Section 13 Troubleshooting
Table 4 Luminometer troubleshooting
Display
Fault
Corrective action
—
Detector overload
Change the measurement range to the 0–2000
range (refer to
section 3.2.3, Set the measurement
range on page 18
).
—
cannot read the display
Change batteries (refer to
section 12.3.1,
Luminometer battery replacement on page 112
).
—
Chemiluminescence line
develops in a bell shaped
curve.
The chemiluminescence reagents are weak and
need to be replaced.
—
Negative percent inhibition
Do the reference measurement again.
Salt could be in the sample or the reagents may be
faulty.
—
—
Change the contrast.
Error
Database is full. No new
measurements can be saved.
Erase all measurements (refer to
section 3.2.2,
Erase the results saved on the luminometer on
page 17
).
01
System RAM test failed
Memory failure. Contact technical support.
02
System EPROM test failed
Memory failure. Contact technical support.
03
LCD display test failed
Continue operation. If fault occurs again, contact
technical support.
04
Non-volatile RAM test failed
Memory failure. Contact technical support.
05
Storage measurements corrupt
All saved measurements will be erased. Continue
operation. If fault occurs again, contact technical
support.
06
Configuration settings corrupt
LCD contrast and measurement range will be
reset. Continue operation. If fault error occurs
again, contact technical support.
07
Usage counter corrupt
Usage counter will be reset to zero. Continue
operation. If fault occurs again, contact technical
support.
08
A/D input failure
Failed measurement. Contact technical support.
09
Reference LED failure
Failed measurement. Contact technical support.
11
Reference LED reading fault
Failed measurement. Contact technical support.
100
Cannot clear measurements
Remove power and then apply power to the
luminometer. If fault occurs again, contact
technical support.
101
Cannot store range settings
Remove power and then apply power to the
luminometer. If fault occurs again, contact
technical support.
Содержание Eclox
Страница 1: ...2886888 Eclox Rapid Response Test Kit User Manual 01 2018 Edition 5 ...
Страница 2: ...Visit us at www hach com ...
Страница 6: ...Table of Contents Table of Contents Page 6 EcloxTOC fm ...
Страница 14: ...14 ...
Страница 24: ...24 ...
Страница 46: ...46 ...
Страница 52: ...52 ...
Страница 110: ...110 ...
Страница 113: ...113 ...
Страница 114: ...Visit us at www hach com ...
Страница 120: ...120 ...
Страница 123: ...123 ...