10.13 Show or send previous results
To show previous results on the luminometer, do the procedure in this section for the
type of procedure done.
To send previous results to a computer:
Note:
At this stage, the results cannot be sent to the LUMISsoft 4.
1.
Do the steps in
Connect the luminometer to a computer on page 23
.
2.
Start LUMISsoft.
3.
In LUMISsoft, select Transfer, Options, Interface Protocol, Connect.
4.
Do the procedure in this section for the type of procedure done.
To send previous results to a printer, do the steps in
Connect the luminometer to a
printer on page 21
and then do the procedure in this section for the type of
procedure done.
10.13.1 Description of screening luminescence results
Reference (non-toxic) luminescent measurements are saved as R1 to Rx. The
counter starts with R1 every time the storage is erased from the luminometer.
Sample luminescent measurements are done after a reference luminescent
measurement is done. Sample luminescent measurements are saved as S1 to Sx.
The luminometer records reference and sample measurements and then calculates
the percent inhibition value for each sample measurement (refer to
Figure 4
).
For example, two different screening luminescence tests have been done. One test
with 3 samples or sample dilutions and one test with two samples or sample
dilutions. The results of the first test are indicated as R1 with S1,S2 and S3. The
results of the second tests are indicated as R2 with S1 and S2.
Figure 4 Example of screening luminescence results
Содержание Eclox
Страница 1: ...2886888 Eclox Rapid Response Test Kit User Manual 01 2018 Edition 5 ...
Страница 2: ...Visit us at www hach com ...
Страница 6: ...Table of Contents Table of Contents Page 6 EcloxTOC fm ...
Страница 14: ...14 ...
Страница 24: ...24 ...
Страница 46: ...46 ...
Страница 52: ...52 ...
Страница 110: ...110 ...
Страница 113: ...113 ...
Страница 114: ...Visit us at www hach com ...
Страница 120: ...120 ...
Страница 123: ...123 ...