32
16.
Close the lid. Push
PROCEED
.
The luminometer
automatically starts
measuring. After four
minutes, the screen timer
shows DONE.
17.
The Inhib% is shown
on the screen. Record the
Inhib% value and graph on
the Test Record Sheet.
Note:
For sea water samples,
the graph may appear higher
than the reference and the
Inhib% may be negative.
18.
Push
PROCEED
to go
back to the Measure Menu.
19.
Remove the cuvette
from the luminometer cell.
Put the solution into the
waste bottle. Put the
cuvette into the waste bag.
20.
Sign the Test Record
Sheet.
Put the sample from the
beaker in the wastewater
drain using local operating
procedures.