80
4.
Put 1 part stock
suspension (S) at
refrigerator temperature
into a clean reaction vessel
using a pipette.
For example:
0.1 mL S + 2.0 mL D
5.
Put the cap on the
reaction vessel and shake
to mix thoroughly.
Put a “1:20” label on the
reaction vessel.
6.
Put 50 parts Diluent
solution (D) at refrigerator
temperature into a clean
reaction vessel using a
pipette.
7.
Put 1 part stock
suspension (S) at
refrigerator temperature
into a clean reaction vessel
using a pipette.
For example:
0.2 mL S + 10 mL D
8.
Put the cap on the
reaction vessel and shake
to mix thoroughly.
Put a “1:50” label on the
reaction vessel.
9.
Put one half of the new,
empty test tubes in Row B
and one half of the test
tubes in Row C of the
LUMIStherm.
Note:
The LUMIStherm
should be operating at 15 °C.