FETCh (READ) commands
FETCh and READ commands are similar. FETCh command is used
to return the last measurement data, and READ command is used
to return the latest measurement data. Therefore, data is returned
after a complete measurement cycle through READ commands and
the implementation efficiency is slightly worse at slow
measurement.
FETCh commands are used to obtain test data. Before using this
command, you need to set the [Result] option to FETCH on the
[SYSTEM CONFIG] page.
:FETCh? ........................................................................... 170
:FETch:FULL .................................................................... 171
:FETCh?
Description
Returns the main test data.
Query Syntax
:FETCh?
Parameter
<R>,<V> Return both the resistance and voltage
value
<R>
Return the resistance value
<V>
Return the resistance value
Example
->:FETC?
->22.005E+0, 3.0
Note
If current LCD screen is not displayed on [MEAS
DISPLAY] or [ENLARGE DISPLAY] page, the LCD
screen will be switched to [MEAS DISPLAY] page
before returning the measurement data when using
this command.
Query
Содержание GBM-3080
Страница 1: ...Battery Meter GBM 3080 3300 USER MANUAL Rev 1 02 ISO 9001 CERTIFIED MANUFACTURER ...
Страница 21: ...GETTING STARTED 21 Carry position ...
Страница 85: ...SYSTEM CONFIGURATION 85 SYSTEM CONFIG MEAS SETUP ...
Страница 87: ...SYSTEM CONFIGURATION 87 Note There is not any option that user can configure on System Information page ...
Страница 103: ...OTHER FUNCTIONS 103 Output to PLC negative common terminal Output to PLC positive common terminal ...
Страница 127: ...COMMAND OVERVIEW 127 Note NULL will be returned if the prompt field is empty ...
Страница 159: ...COMMAND OVERVIEW 159 Note Process capability index deviation Process capability index offset ...
Страница 174: ...GBM 3080 3300 User Manual 174 Parameter None Current file File No 0 9 Specified file Example FILE DEL FILE DEL 1 ...
Страница 183: ...APPENDIX 183 Dimensions ...