GBM-3080/3300 User Manual
164
:SYSTem:CODE
Description
Sets or returns the state of the error code function.
Syntax
:SYSTem:CODE {OFF/0, ON/1}
Query Syntax
:SYSTem:CODE?
Parameter/
Return parameter
0
Turn the error code function off.
1
Turn the error code function on.
Example
->:SYST:CODE ON
->:SYST:CODE?
->ON
Note
If command “:SYST:CODE” is enabled, the
instrument will return an error code each time it
receives an instruction. If command “:SYST:CODE
is disabled, user can obtain error code through
command “ERR?”
E00: No error.
E01: Bad command.
E02: Parameter error.
E03: Missing parameter.
E04: Buffer overruns.
E05: Syntax error.
E06: Invalid separator.
E07: Invalid multiplier.
E08: Numeric data error.
E09: Value too long
E10: Invalid command
E11: Unknown error
:SYSTem:BEEPer
Description
Sets or returns the key click beeper state.
Syntax
:SYSTem:BEEPer {OFF/0, ON/1}
Query Syntax
:SYSTem:BEEPer?
Set
Query
Set
Query
Содержание GBM-3080
Страница 1: ...Battery Meter GBM 3080 3300 USER MANUAL Rev 1 02 ISO 9001 CERTIFIED MANUFACTURER ...
Страница 21: ...GETTING STARTED 21 Carry position ...
Страница 85: ...SYSTEM CONFIGURATION 85 SYSTEM CONFIG MEAS SETUP ...
Страница 87: ...SYSTEM CONFIGURATION 87 Note There is not any option that user can configure on System Information page ...
Страница 103: ...OTHER FUNCTIONS 103 Output to PLC negative common terminal Output to PLC positive common terminal ...
Страница 127: ...COMMAND OVERVIEW 127 Note NULL will be returned if the prompt field is empty ...
Страница 159: ...COMMAND OVERVIEW 159 Note Process capability index deviation Process capability index offset ...
Страница 174: ...GBM 3080 3300 User Manual 174 Parameter None Current file File No 0 9 Specified file Example FILE DEL FILE DEL 1 ...
Страница 183: ...APPENDIX 183 Dimensions ...