GBM-3080/3300 User Manual
152
Note
The number of decimal digits and unit are related to
the current range no. Please refer to the following
table for their relationship.
Range no.
Number of decimal digits
Unit Description
0(8V)
5
V
100000=1.00000V
1(80V)
4
V
100000=10.0000V
2(300V
)
3
V
100000=100.000V
:CALCulate:LIMit:VOLTage:PERCent
Description
Sets and returns limit in percentage for voltage
comparator.
Syntax
:CALCulate:LIMit:VOLTage:PERCent<float> (0.0 ~
100.0)
Query Syntax
:CALCulate:LIMit:VOLTage:PERCent?
Parameter /
Return parameter
(0.0 ~ 100.0) It is a floating-point number without
positive and negative symbols.
Example
->:CALC:LIM:VOLT:PERC 1.1
->:CALC:LIM:VOLT:PERC?
->1.100
Note
:CALCulate:LIMit:ABS
Description
Sets absolute deviation comparison for voltage.
Syntax
:CALCulate:LIMit:ABS {ON/1, OFF/0}
Query Syntax
:CALCulate:LIMit:ABS?
Parameter /
Return parameter
ON
Sets the comparator for voltage to absolute
deviation comparison.
OFF
Sets the comparator for voltage to relative
deviation (percentage) comparison.
Example
->:CALC:LIM:ABS ON
->:CALC:LIM:ABS?
->ON
Set
Query
Set
Query
Содержание GBM-3080
Страница 1: ...Battery Meter GBM 3080 3300 USER MANUAL Rev 1 02 ISO 9001 CERTIFIED MANUFACTURER ...
Страница 21: ...GETTING STARTED 21 Carry position ...
Страница 85: ...SYSTEM CONFIGURATION 85 SYSTEM CONFIG MEAS SETUP ...
Страница 87: ...SYSTEM CONFIGURATION 87 Note There is not any option that user can configure on System Information page ...
Страница 103: ...OTHER FUNCTIONS 103 Output to PLC negative common terminal Output to PLC positive common terminal ...
Страница 127: ...COMMAND OVERVIEW 127 Note NULL will be returned if the prompt field is empty ...
Страница 159: ...COMMAND OVERVIEW 159 Note Process capability index deviation Process capability index offset ...
Страница 174: ...GBM 3080 3300 User Manual 174 Parameter None Current file File No 0 9 Specified file Example FILE DEL FILE DEL 1 ...
Страница 183: ...APPENDIX 183 Dimensions ...