GBM-3080/3300 User Manual
36
Setting the output current mode
When several identical devices measure in parallel, the measured
signals will interfere with each other, causing the measured value
to change suddenly. To prevent measurement error, change the
current output mode to
PAUSE
which will turn off the current
source after the test is completed to minimize the interference of
multiple devices.
Steps
1.
Press the
Setup
button to enter
[MEAS SETUP] page.
2.
Use arrow keys to move cursor and
select
CURRENT
on the [MEAS
SETUP] page.
3.
Use option key on the right of the LCD screen
to select an available parameter.
Available
parameter
CONTINUOUS Output current continuously.
PAUSE
Output current only during
measurement and turn off signal
source after measurement is
completed.
Содержание GBM-3080
Страница 1: ...Battery Meter GBM 3080 3300 USER MANUAL Rev 1 02 ISO 9001 CERTIFIED MANUFACTURER ...
Страница 21: ...GETTING STARTED 21 Carry position ...
Страница 85: ...SYSTEM CONFIGURATION 85 SYSTEM CONFIG MEAS SETUP ...
Страница 87: ...SYSTEM CONFIGURATION 87 Note There is not any option that user can configure on System Information page ...
Страница 103: ...OTHER FUNCTIONS 103 Output to PLC negative common terminal Output to PLC positive common terminal ...
Страница 127: ...COMMAND OVERVIEW 127 Note NULL will be returned if the prompt field is empty ...
Страница 159: ...COMMAND OVERVIEW 159 Note Process capability index deviation Process capability index offset ...
Страница 174: ...GBM 3080 3300 User Manual 174 Parameter None Current file File No 0 9 Specified file Example FILE DEL FILE DEL 1 ...
Страница 183: ...APPENDIX 183 Dimensions ...