APPENDIX
181
Interface
RS232C
USB Port
Handler
Programming
language
SCPI
Accessibility Keypad lock
AC Resistance
Range
No.
Range
Maximum
Displayed Values
Resolution
Measured
current
0
3m
Ω
3.1000m
0.1
μΩ
100mA
1
30m
Ω
31.000m
1
μΩ
100mA
2
300m
Ω
310.00m
10
μΩ
10mA
3
3
Ω
3.1000
100
μΩ
1mA
4
30
Ω
31.000
1m
Ω
100
μ
A
5
300
Ω
310.00
10m
Ω
10
μ
A
6
3k
Ω
3200.0
100m
Ω
10
μ
A
Range
No.
Accuracy
Temperature
coefficient
Slow
Medium
Fast
Ex. Fast
0
±0.5%rdg
±10 dgt
±0.5%rdg
±15 dgt
±0.5%rdg
±20 dgt
±0.5%rdg
±40 dgt
(±0.05% rdg
±1dgt)/˚C
1~6
±0.5% rdg
±5 dgt
±0.5% rdg
±7 dgt
±0.5% rdg
±7 dgt
±1% rdg
±8 dgt
(±0.05% rdg
± 0.5 dgt)/˚C
DC Voltage
Range
No.
Range
Maximum Displayed
Values
Resolution
0
8V
±
8.08000
10
μV
1
80V
±
80.8000
100
μV
2
300V (For GBM-3300 only)
±
303.000
1mV
Range
No.
Accuracy
Temperature
coefficient
Slow
Medium
Fast
Ex. Fast
0~2
±
0.01%
rdg
±
3dgt
±
0.01%
rdg
±
5dgt
±
0.05%
rdg
±
5dgt
±
0.1% rdg
±
6dgt
(±0.001% rdg
±0.3dgt)/˚C
Содержание GBM-3080
Страница 1: ...Battery Meter GBM 3080 3300 USER MANUAL Rev 1 02 ISO 9001 CERTIFIED MANUFACTURER ...
Страница 21: ...GETTING STARTED 21 Carry position ...
Страница 85: ...SYSTEM CONFIGURATION 85 SYSTEM CONFIG MEAS SETUP ...
Страница 87: ...SYSTEM CONFIGURATION 87 Note There is not any option that user can configure on System Information page ...
Страница 103: ...OTHER FUNCTIONS 103 Output to PLC negative common terminal Output to PLC positive common terminal ...
Страница 127: ...COMMAND OVERVIEW 127 Note NULL will be returned if the prompt field is empty ...
Страница 159: ...COMMAND OVERVIEW 159 Note Process capability index deviation Process capability index offset ...
Страница 174: ...GBM 3080 3300 User Manual 174 Parameter None Current file File No 0 9 Specified file Example FILE DEL FILE DEL 1 ...
Страница 183: ...APPENDIX 183 Dimensions ...