MEASUREMENT
53
DECL-
Decrease the range number and set to
hold range.
Range for resistance
The GBM-3000 series has seven ranges for resistance with varying
ranges for each range as follows:
Range no.
Range name
Range
0
3m
Ω
0.0000m
Ω
~ 3.1000m
Ω
1
30m
Ω
0.000m
~ 31.000m
2
300m
Ω
0.00m
Ω
~ 310.00m
Ω
3
3
Ω
0.0000
~ 3.1000
4
30
Ω
0.000
~ 31.000
5
300
Ω
0.00
~ 310.00
6
3k
Ω
0.0
~ 3200.0
Range for voltage
The GBM-3000 series has three ranges for voltage with varying
ranges for each range as follows:
Range no.
Range name
Range
0
8V
0.00000V~8.08000V
1
80V
0.0000V~80.8000V
2
300V
0.000V~303.000V(For GBM-3300 only)
Abnormal value description
Display on the screen
Description
Unable to determine
Measured value is higher than the measuring
range
Measured value is lower than the measuring
range
Содержание GBM-3080
Страница 1: ...Battery Meter GBM 3080 3300 USER MANUAL Rev 1 02 ISO 9001 CERTIFIED MANUFACTURER ...
Страница 21: ...GETTING STARTED 21 Carry position ...
Страница 85: ...SYSTEM CONFIGURATION 85 SYSTEM CONFIG MEAS SETUP ...
Страница 87: ...SYSTEM CONFIGURATION 87 Note There is not any option that user can configure on System Information page ...
Страница 103: ...OTHER FUNCTIONS 103 Output to PLC negative common terminal Output to PLC positive common terminal ...
Страница 127: ...COMMAND OVERVIEW 127 Note NULL will be returned if the prompt field is empty ...
Страница 159: ...COMMAND OVERVIEW 159 Note Process capability index deviation Process capability index offset ...
Страница 174: ...GBM 3080 3300 User Manual 174 Parameter None Current file File No 0 9 Specified file Example FILE DEL FILE DEL 1 ...
Страница 183: ...APPENDIX 183 Dimensions ...