DATA LOGGING AND STATISTICS
71
Process Capability Index (Offset) CpK
Note
N stands for valid data which means that overflow
and open values are excluding. As long as the
digital data can be displayed on the screen, it is
regarded as valid value.
The Hi and Lo variables in the Cp and CpK
formulas are the actual upper and lower limits of
the comparator. In the PER and ABS comparison
modes, the actual value is converted from the
nominal value. This value will be taken into
operation regardless of whether the comparator is
enabled or not.
When sample standard deviation
σ
n-1 = 0, Cp =
99.99, CpK = 99.99
When CpK<0, Cpk=0.
The following SCPI query commands are used to obtain related
information.
Number of
statistics
CALCulate:STATistic:RESistance:NUMBer?
CALCulate:STATistic:VOLTage:NUMBer?
For detailed, please refer to page 153 and 156.
The average of
statistics
CALCulate:STATistic:RESistance:MEAN?
CALCulate:STATistic:VOLTage:MEAN?
For detailed, please refer to page 154 and 156.
The maximum
value of statistics
CALCulate:STATistic:RESistance:MAXimum?
CALCulate:STATistic:VOLTage:MAXimum?
For detailed, please refer to page 154 and 157.
Содержание GBM-3080
Страница 1: ...Battery Meter GBM 3080 3300 USER MANUAL Rev 1 02 ISO 9001 CERTIFIED MANUFACTURER ...
Страница 21: ...GETTING STARTED 21 Carry position ...
Страница 85: ...SYSTEM CONFIGURATION 85 SYSTEM CONFIG MEAS SETUP ...
Страница 87: ...SYSTEM CONFIGURATION 87 Note There is not any option that user can configure on System Information page ...
Страница 103: ...OTHER FUNCTIONS 103 Output to PLC negative common terminal Output to PLC positive common terminal ...
Страница 127: ...COMMAND OVERVIEW 127 Note NULL will be returned if the prompt field is empty ...
Страница 159: ...COMMAND OVERVIEW 159 Note Process capability index deviation Process capability index offset ...
Страница 174: ...GBM 3080 3300 User Manual 174 Parameter None Current file File No 0 9 Specified file Example FILE DEL FILE DEL 1 ...
Страница 183: ...APPENDIX 183 Dimensions ...