SETTING UP
35
Available
parameter
ON
Enable the self-calibration function. The
device will perform a self-calibration every
30 minutes.
After the self-calibration, a message "Self-
Calibration was successful" will display on
the message column below the LCD to
indicate that self-calibration has been
completed.
OFF
Disable the self-calibration function.
Note
When self-calibration is performed, the measurement
will pause briefly to respond to the self-calibration.
A self-calibration takes 40ms. It is necessary to
disable the self-calibration function when performing
high-speed measurement and use external I/O
control cord for self-calibration.
To ensure accuracy, the device performs a self-
calibration each time when it is turned on.
In addition to regular self-calibration, you also can
perform a self-calibration by using
1. The SELF.CAL signal of HANDLER.
2. The communication command [SYST: CALibration].
Содержание GBM-3080
Страница 1: ...Battery Meter GBM 3080 3300 USER MANUAL Rev 1 02 ISO 9001 CERTIFIED MANUFACTURER ...
Страница 21: ...GETTING STARTED 21 Carry position ...
Страница 85: ...SYSTEM CONFIGURATION 85 SYSTEM CONFIG MEAS SETUP ...
Страница 87: ...SYSTEM CONFIGURATION 87 Note There is not any option that user can configure on System Information page ...
Страница 103: ...OTHER FUNCTIONS 103 Output to PLC negative common terminal Output to PLC positive common terminal ...
Страница 127: ...COMMAND OVERVIEW 127 Note NULL will be returned if the prompt field is empty ...
Страница 159: ...COMMAND OVERVIEW 159 Note Process capability index deviation Process capability index offset ...
Страница 174: ...GBM 3080 3300 User Manual 174 Parameter None Current file File No 0 9 Specified file Example FILE DEL FILE DEL 1 ...
Страница 183: ...APPENDIX 183 Dimensions ...