COMMAND OVERVIEW
161
This command is valid only when Data logging
function is enabled. Confirm the status of [DATA
logging] on [SYSTEM CONFIG]Page.
If the current page is not on [MEAS DISPLAY] page,
this command will automatically switch [MEAS
DISPLAY] page.
:LOGger:SIZE
Description
Sets or returns buffer size for data recording.
Syntax
:LOGger:SIZE{<1~10000>, max}
Query Syntax
:LOGger:SIZE?
Parameter /
Return parameter
<1~10000> Integer. If the number is less than 1, it
will be set to 1 automatically.
Max
Set the buffer to 10000
Example
->:LOG:SIZE 100
->:LOG:SIZE?
->100
:LOGger:COUNt
Description
Queries the total number of recorded data in
buffer
Query Syntax
:LOGger:COUNt?
Parameter /
Return parameter
0~10000
If return value is zero, it means buffer is
empty.
Example
->:LOG:COUN?
->10
:LOGger:DATA?
Description
Queries the value of data in buffer
Query Syntax
:LOGger:DATA?
Set
Query
Set
Query
Set
Query
Содержание GBM-3080
Страница 1: ...Battery Meter GBM 3080 3300 USER MANUAL Rev 1 02 ISO 9001 CERTIFIED MANUFACTURER ...
Страница 21: ...GETTING STARTED 21 Carry position ...
Страница 85: ...SYSTEM CONFIGURATION 85 SYSTEM CONFIG MEAS SETUP ...
Страница 87: ...SYSTEM CONFIGURATION 87 Note There is not any option that user can configure on System Information page ...
Страница 103: ...OTHER FUNCTIONS 103 Output to PLC negative common terminal Output to PLC positive common terminal ...
Страница 127: ...COMMAND OVERVIEW 127 Note NULL will be returned if the prompt field is empty ...
Страница 159: ...COMMAND OVERVIEW 159 Note Process capability index deviation Process capability index offset ...
Страница 174: ...GBM 3080 3300 User Manual 174 Parameter None Current file File No 0 9 Specified file Example FILE DEL FILE DEL 1 ...
Страница 183: ...APPENDIX 183 Dimensions ...