GBM-3080/3300 User Manual
146
OFF Disable both the comparators for voltage and
for resistance.
Example
->:CALC:LIM:STAT OFF
->:CALC:LIM:STAT?
->OFF
Note
:CALCulate:LiMit:STATe command enables or disables
the comparators for voltage and for resistance
simultaneously.
If a single comparator for resistance or voltage needs
to be used, use the :RES:LMT:STATe
and :VOLT:LMT:STATe commands.
:CALCulate:LIMit:BEEPer
Description
Sets the comparator buzzer.
Syntax
:CALCulate:LIMit:BEEPer{0/OFF, HL/NG/FAIL,
IN/OK/PASS}
Query Syntax
:CALCulate:LIMit:PEEPer?
Return parameter OFF
Disable the comparator buzzer.
HL
Buzzer sounds when testing is failed
IN
Buzzer sounds when testing is passed
Example
->:CALC:LIM:BEEP HL
->:CALC:LIM:BEEP?
->HL
:CALCulate:LIMit:RESistance:MODE
Description
Sets and returns the comparator mode for
resistance.
Syntax
:CALCulate:LIMit:RESistance:MODE{HL, REF, ABS}
Query Syntax
:CALCulate:LIMit:RESistance:MODE?
Parameter /
Return parameter
HL
Upper and lower limits of direct readout
comparison(SEQ)
Set
Query
Set
Query
Содержание GBM-3080
Страница 1: ...Battery Meter GBM 3080 3300 USER MANUAL Rev 1 02 ISO 9001 CERTIFIED MANUFACTURER ...
Страница 21: ...GETTING STARTED 21 Carry position ...
Страница 85: ...SYSTEM CONFIGURATION 85 SYSTEM CONFIG MEAS SETUP ...
Страница 87: ...SYSTEM CONFIGURATION 87 Note There is not any option that user can configure on System Information page ...
Страница 103: ...OTHER FUNCTIONS 103 Output to PLC negative common terminal Output to PLC positive common terminal ...
Страница 127: ...COMMAND OVERVIEW 127 Note NULL will be returned if the prompt field is empty ...
Страница 159: ...COMMAND OVERVIEW 159 Note Process capability index deviation Process capability index offset ...
Страница 174: ...GBM 3080 3300 User Manual 174 Parameter None Current file File No 0 9 Specified file Example FILE DEL FILE DEL 1 ...
Страница 183: ...APPENDIX 183 Dimensions ...