OTHER FUNCTIONS
89
Offset adjustment by short test
Before performing measure, be sure to perform offset for short test
to remove stray resistance and bias due to the test leads and
external environmental conditions.
In order to meet the specification of the device, offset for short test
must be performed.
Note
In addition to perform user calibration through
SHORT TEST page, user can perform user calibration
through other methods.
1. by using the nSHORT signal of HANDLER.
2. by using the communication command [:ADJust].
Since generated voltage signal will be very small (several mV at
maximum) after the test current flows through the resistor in the
condition of very small measuring resistance (3mΩ and 30mΩ
ranges), the location, length and shape of the test leads may have
influences on measuring.
In general, the location where measurement is performed is also the
location where short circuit resetting is performed.
Example
Be sure to keep consistent of location when
performing short circuit offsetting, especially
when the range is 30m
Ω
or below in
measurement.
Содержание GBM-3080
Страница 1: ...Battery Meter GBM 3080 3300 USER MANUAL Rev 1 02 ISO 9001 CERTIFIED MANUFACTURER ...
Страница 21: ...GETTING STARTED 21 Carry position ...
Страница 85: ...SYSTEM CONFIGURATION 85 SYSTEM CONFIG MEAS SETUP ...
Страница 87: ...SYSTEM CONFIGURATION 87 Note There is not any option that user can configure on System Information page ...
Страница 103: ...OTHER FUNCTIONS 103 Output to PLC negative common terminal Output to PLC positive common terminal ...
Страница 127: ...COMMAND OVERVIEW 127 Note NULL will be returned if the prompt field is empty ...
Страница 159: ...COMMAND OVERVIEW 159 Note Process capability index deviation Process capability index offset ...
Страница 174: ...GBM 3080 3300 User Manual 174 Parameter None Current file File No 0 9 Specified file Example FILE DEL FILE DEL 1 ...
Страница 183: ...APPENDIX 183 Dimensions ...