APS-7000 Programming Manual
98
Syntax
[:SOURce]:FUNCtion:CSINe:TYPE
{CLP1|CLP2|CLP3,CFACtor|CLIP|SDIP|STAircase|
TRIangle}
Query Syntax
[:SOURce]:FUNCtion:CSINe:TYPE? {CLP1|CLP2|CLP3}
Parameter/Return
parameter
CLP1
Save slot 1
CLP2
Save slot 2
CLP3
Save slot 3
CFACtor
Set the type as a crest factor waveform.
CLIP
Set the type as a clipped sine waveform.
SDIP
Set the type as a surge/dip waveform.
STAircase
Set the type as a staircase waveform.
Triangle
Set the type as a triangle waveform.
Example
:FUNC:CSIN:TYPE CLP1,CFACtor
Sets the arbitrary waveform type as a crest factor
waveform.
[:SOURce]:FUNCtion[:SHAPe][:IMMediate]
Set
Query
Description
Loads the selected arbitrary waveform into the
power supply. The selected waveform will be
output when the output is turned on.
Note:
If the CLP1|2|3 arbitrary waveforms are to be
used, they must first be setup before they can be
output. Use the [:SOURce]:FUNCtion:CSINe:TYPE
command to set the CLP1|2|3 save slots.
Syntax
[:SOURce]:FUNCtion[:SHAPe][:IMMediate]
{SIN|ARB1|ARB2|ARB3|ARB4|ARB5|ARB6|ARB7|ARB8|
ARB9|ARB10|ARB11|ARB12|ARB13|ARB14|ARB15|ARB
16|ARB17|ARB18|ARB19|ARB20|ARB21|ARB22|CLP1|C
LP2|CLP3}
Query Syntax
[:SOURce]:FUNCtion[:SHAPe][:IMMediate]?
Parameter/Return
parameter
SIN
Sine waveform
ARB1 ~
ARB22
Preset ARB waveforms, number 1 ~ 22.
See the user manual for details.
CLP1
Save slot 1
CLP2
Save slot 2
CLP3
Save slot 3
All manuals and user guides at all-guides.com