REMOTE CONTROL
109
Parameter/Return
parameter
<NR2>
OFF Phs (Stop phase)
MINimum
0
MAXimum
359
Example
:SIM:INIT:PHAS:STOP 0
Sets OFF Phs to 0.
[:SOURce]:SIMulation:INITial:VOLTage
Set
Query
Description
Sets or queries the Vset parameter of the abnormal
step for the Simulation mode.
Syntax
[:SOURce]:SIMulation:INITial:VOLTage
{<NR2>|MINimum|MAXimum}
Query Syntax
[:SOURce]:SIMulation:INITial:VOLTage?
[MINimum|MAXimum]
Parameter/Return
parameter
<NR2>
Voltage of the initial step.
MINimum
Minimum settable voltage
MAXimum
Maximum settable voltage
Example
:SIM:INIT:VOLT MAX
Sets the initial step voltage to the maximum.
[:SOURce]:SIMulation:NORMal<1|2>
:CODE
Set
Query
Description
Sets the external trigger output for the normal 1 or
normal 2 step parameter. This option is only
applicable when in the Simulation mode.
Syntax
[:SOURce]:SIMulation:NORMal<1|2>:CODE
{<NR1>|MINimum|MAXimum}
Query Syntax
[:SOURce]:SIMulation:NORMal<1|2>:CODE?
[MINimum|MAXimum]
Parameter/Return
parameter
<NR1>
HI=1, LO=0
MINimum
LO, 0
MAXimum
HI, 1
Example
SIM:NORM1:CODE 1