REMOTE CONTROL
55
Initiate Commands
:INITiate[:IMMediate]:NAME ................................. 55
:INITiate[:IMMediate][:TRANsient] ....................... 55
:INITiate[:IMMediate]:NAME
Set
Description
The INITiate command starts the TRANsient,
OUTPut, MEMory or SDIP (surge/dip) trigger.
Syntax
:INITiate[:IMMediate]:NAME
{TRANsient|OUTPut|MEMory|SDIP}
Parameter
TRANsient
Starts the TRANSient trigger.
OUTPut
Starts the OUTput trigger.
MEMory
Starts the MEMory trigger.
SDIP
Starts the (SDIP) Surge/Dip trigger.
Example
:INIT:NAME TRAN
Starts the transient trigger.
:INITiate[:IMMediate][:TRANsient]
Set
Description
This command controls the enabling of output
triggers. When a trigger is enabled, a trigger causes
the specified action to occur. If the trigger system
is not enabled, all triggers are ignored.
Syntax
:INITiate[:IMMediate][:TRANsient]
Example
:INIT