REMOTE CONTROL
95
Example
:FREQ
60.00
Sets the frequency of 60Hz.
[:SOURce]:FUNCtion:CSINe:CFACtor
Set
Query
Description
Sets or queries the crest factor setting for the
waveform.
Note:
The :SOURce:FUNCtion:CSINe:TYPE command
must first be used to set the save slot number
(CLP1|2|3) and CFACtor as the waveform type
before this command is executed.
Syntax
[:SOURce]:FUNCtion:CSINe:CFACtor
{CLP1|CLP2|CLP3,<NR2>|MINimum|MAXimum}
Query Syntax
[:SOURce]:FUNCtion:CSINe:CFACtor?
{CLP1|CLP2|CLP3[,MINimum|MAXimum]}
Parameter/Return
parameter
<NR2>
Crest factor
CLP1
Save slot 1
CLP2
Save slot 2
CLP3
Save slot 3
MIN
1.4
MAX
10.0
Example
:FUNC:CSIN:CFAC CLP1,2.0
Sets the crest factor to 2.0.
[:SOURce]:FUNCtion:CSINe:CLIP
Set
Query
Description
Sets or queries the CLIP waveform settings.
Note:
The :SOURce:FUNCtion:CSINe:TYPE command
must first be used to set the save slot number
(CLP1|2|3) and CLIP as the waveform type before
this command is executed.
Syntax
[:SOURce]:FUNCtion:CSINe:CLIP
{CLP1|CLP2|CLP3,<NR2>|MINimum|MAXimum}