![GW Instek APS-7100 Скачать руководство пользователя страница 52](http://html1.mh-extra.com/html/gw-instek/aps-7100/aps-7100_programming-manual_2279972052.webp)
APS-7000 Programming Manual
52
Data/Trace Commands
Note
The DATA and the TRACe node for the following
commands are functionally equivalent.
:DATA|TRACe:SEQuence:CLEar .......................... 52
:DATA|TRACe:SEQuence:RECall ......................... 52
:DATA|TRACe:SEQuence:STORe ......................... 53
:DATA|TRACe:SIMulation:CLEar ......................... 53
:DATA|TRACe:SIMulation:RECall ........................ 53
:DATA|TRACe:SIMulation:STORe ........................ 54
:DATA|TRACe:SEQuence:CLEar
Set
Description
Clears the sequence data for the selected save
memory (Seq0 ~ Seq9).
Syntax
:DATA|TRACe:SEQuence:CLEar
{<NR1>|MINimum|MAXimum}
Parameter
<NR1>
0~9
MIN
0
MAX
9
Example
:DATA:SEQ:CLE 1
Clears the sequence data from Seq1.
:DATA|TRACe:SEQuence:RECall
Set
Description
Loads the sequence data. This command is the
equivalent to recalling a sequence memory in the
Sequence mode.
Syntax
:DATA|TRACe:SEQuence:RECall
{<NR1>|MINimum|MAXimum}
Parameter
<NR1>
0~9 (Seq0 ~ Seq9).
MIN
0
MAX
9
All manuals and user guides at all-guides.com