Preliminary Device Description HG G-19600ZA | English, Revision 02 | Date: 13.03.2018
15
Commissioning – Chapter 4
Figure 8
Commissioning: Menus of the service program
Meaning of the highlighted parameters:
•
track polarity
in
settings
menu: Polarity of the magnetic tape
•
sensitivity
in
settings
menu: Sensitivity of the detection of a magnetic track, threshold
value in milliTesla (mT)
•
sen. two tracks
in
settings
menu: Sensitivity of the detection of two magnetic tracks.
For applications without junctions set to maximum value
4
.
•
sum of flux density
in
raw data menu
: Flux density of a magnetic tape underneath the
sensor
Tip
When using one of the magnetic tapes listed in section 3.1.1 on page 10 (HW MAT00003,
HW MAT00008 and HW MAT00009) the threshold sen. two tracks for two magnetic tracks
can be set according to the following table:
Reading height [mm]
50
60
70
80
90
100
sen. two tracks [mT]
0.8
0.64
0.5
0.4
0.32
0.275
Table 3
Thresholds for two magnetic tracks with standard magnetic tapes
HG19600 Serialnumber: 999999 Firmwareversion: V0.45
system monitor
__ __
| | | |
---------------------------------
| HG 19600 O O X O X |
| |
---------------------------------
X
next turn: STRAIGHT
position track: 7.500
(L) left
(R) right
(S) straight
error: 0x00 (D) digital input
(1) CSV Output
(2) settings
(3) testmode
(4) raw data
Status: DIN:00 DOUT:0 AOUT1(X): 7.5V AOUT2(Z): 7.6V
HG 19600 Serialnumber: 999999 Firmwareversion: V0.45
settings
Range:
(1) track polarity : north
(3) sensitivity : 0.200 [mT] 0.01 .. 2
(4) sen. two tracks: 0.800 0.1 .. 4
AOUT1 (X)
(5) amplitude : 10.00 [V] 2.5 .. (10 - offset)
(6) offset : 0.00 [V] 0 .. (10 - amplitude)
(7) invert output : NO
(H) hold analog output when detect is lost: NO
AOUT2 (Z)
(8) amplitude : 10.00 [V] 0 .. 10
(9) multiplier : 1.00 0 .. 10
(C) calibration
(D) load default settings
(F) firmware update
(ESC) QUIT
HG 19600 Serialnumber: 999999 Firmwareversion: V0.45
RAW data monitor
__ __
| | | |
---------------------------------
| HG 19600 O O O O X |
| |
---------------------------------
system 3 system 2 system 1
X -0.015 [mT] 0.011 [mT] 0.028 [mT]
Y -0.068 [mT] -0.041 [mT] -0.060 [mT]
Z 0.067 [mT] 0.066 [mT] 0.040 [mT]
track No No No
detect
sum of flux density:
0.231
(ESC) QUIT
Status: DIN:00 DOUT:0 AOUT1(X): 0.0V AOUT2(Z): 7.2V
Key 2
Key 4
Back with ESC
Back with ESC
Main menu