Preliminary Device Description HG G-19600ZA | English, Revision 02 | Date: 13.03.2018
27
Software – Chapter 6
6.4.3 (C) calibration settings
By using the key
in the menu
settings
the
calibration settings
are opened.
Figure 20
Menu calibration settings
In the middle part of the menu the set magnetic field strength for the compensated mag-
netic field are listed. Those values are subtracted from the actual detected values.
–
calibration of the track detection, see section 4.3 on page 18.
–
calibration for a static magnetic field, see section 4.2 on page 16.
–
deletion of a calibration for a static magnetic field, see section 4.2 on page 16.
HG19600 Serialnumber: 999999 Firmwareversion: V0.45
calibration settings
compensated magnetic field
system 3
system 2
system1
0.000 [mT]
0.000 [mT]
0.000 [mT]
0.000 [mT]
0.000 [mT]
0.000 [mT]
0.000 [mT]
0.000 [mT]
0.000 [mT]
(1) calibrate for track detection
(2) calibrate for static magnetic field
(3) delete static magnetic field detection
(ESC) QUIT