– the USB connection: The interface cable must not be
removed.
– the test instrument from supply power.
Wait an additional 5 seconds AFTER the transfer dialog
closes before unplugging the USB connection or turning
off the test instrument.
19.3
Backup and Restore via USB Flash Drive
Structures created and measurement data saved at the test
instrument can be backed up to and restored from a USB flash
drive.
A USB flash drive must be connected to the test instrument to
this end. The respective functions are grayed out at the display if
this is not the case.
Backup
➭
Connect a USB flash drive to the test instrument (see section
10.1 “Use of USB Storage Devices”).
➭
Select the
Backup
function under Setup 1/3 > Database 1/2.
The test instrument creates a backup file on the USB flash
drive directly in the root directory. The backup files on the USB
flash drive are named by means of a time stamp (file exten-
sion: .etcbak).
Attention!
!
The memory structure in the test instrument or in the
connected USB flash drive may be destroyed if data
transmission is interrupted.
The transfer dialog is shown in the test instrument’s dis-
play for as long as data is being transferred. During the
transfer procedure, do not under any circumstances dis-
connect:
– the USB connection: The USB flash drive may not be
removed.
– the test instrument from supply power.
Wait an additional 5 seconds AFTER the transfer dialog
closes before removing the USB flash drive or turning off
the test instrument.
Restore
Attention!
!
Existing data in the test instrument is overwritten.
If applicable, create a backup copy first (see above).
Attention!
!
An existing backup can only be restored to a test instru-
ment with the same or a higher firmware/software ver-
sion as the test instrument from which the backup was
created.
➭
Connect a USB flash drive to the test instrument, to which the
test structure has been saved (see section 10.1 “Use of USB
Storage Devices”).
➭
Select the
Restore
function under Setup > Database 2/2 at the
test instrument.
The data are restored.
Attention!
!
The memory structure in the test instrument or in the
connected USB flash drive may be destroyed if data
transmission is interrupted.
The transfer dialog is shown in the test instrument’s dis-
play for as long as data is being transferred. During the
transfer procedure, do not under any circumstances dis-
connect:
– the USB connection: The USB flash drive may not be
removed.
– the test instrument from supply power.
Wait an additional 5 seconds AFTER the transfer dialog
closes before removing the USB flash drive or turning off
the test instrument.