Gossen Metrawatt GmbH
105
Error while writing the “.secu” file to the
USB flash drive
There’s not (no longer) enough avail-
able memory space on the storage
medium.
–
In particular in the case of FAT16
formatted USB flash drives: too
many files on the USB flash drive
–
Current consumption of the utilized
USB flash drive exceeds 500 mA.
–
If the data on the USB flash drive
are corrupt (e.g. due to an error
during data export to the USB flash
drive).
Make sure that at least 100 MB is
available on the USB flash drive and
delete any data files which are no lon-
ger needed.
➭
If the problem persists, save the data
from the USB flash drive to another
storage device and reformat the USB
flash drive (FAT32).
➭
Only use USB flash drives with cur-
rent consumption of less than
500 mA in combination with the
test instrument.
➭
Make sure that the USB flash drive
isn’t disconnected or moved until
the entire data export process has
been completed.
➭
If none of these measures results in
improvement, replace the USB
flash drive. A list of tested USB
flash drives is included in section
13.1.
Error while writing the data backup file to
the USB flash drive
There’s not (no longer) enough avail-
able memory space on the storage
medium.
–
In particular in the case of FAT16
formatted USB flash drives: too
many files on the USB flash drive
–
Current consumption of the utilized
USB flash drive exceeds 500 mA.
–
The USB flash drive was discon-
nected during data import.
–
The USB flash drive is defective or
incompatible with the test instru-
ment.
➭
Make sure that a at least 100 MB is
available on the USB flash drive
and delete any data files which are
no longer required.
➭
If the problem persists, save the
data from the USB flash drive to
another storage device and refor-
mat the USB flash drive (FAT32).
➭
Only use USB flash drives with cur-
rent consumption of less than
500 mA in combination with the
test instrument.
➭
Make sure that the USB flash drive
isn’t disconnected or moved until
the entire data backup process has
been completed.
➭
If none of these measures results in
improvement, replace the USB
flash drive. A list of tested USB
flash drives is included in section
13.1.
Moving of an object has failed
Moving a test object would lead to a
test object ID conflict. The test object
ID already exists for this customer.
➭
Delete the object with duplicate
test object ID.
➭
Select another customer as a relo-
cation target.
Error Message
Possible Causes
Corrective Measures