Gossen Metrawatt GmbH
General Application Errors
–
The inspector to be deleted is cur-
rently selected and thus cannot be
deleted!
➭
Activate a different inspector before
deleting.
–
The 25 A measurement takes too
long.
or
–
The 25 A measurement has been
executed too often (without
pauses).
➭
Wait until the test instrument has
cooled down and then restart the
measurement.
Database Processing Error
–
One of the fields was filled in with
invalid content while processing an
existing database object.
➭
Please be certain to complete all
mandatory fields (identified in red).
➭
If necessary, check your entries to
the fields for invalid special charac-
ters.
–
The test object ID field was not
filled in while creating a new device
or new ME equipment.
➭
Fill in the test object ID field.
–
There’s already an object with the
same test object ID under the
“Customer” database object.
An incorrect barcode has been
selected.
➭
Assign another test object ID.
Error Message
Possible Causes
Corrective Measures