Gossen Metrawatt GmbH
93
17.9
Remote Control –
Automated Test Sequence Control with IZYTRONIQ Soft-
ware
An IZY remote sequence can be set up in IZYTRONIQ (not to be
confused with a “test instrument sequence”). This is a user-
defined test sequence that can run via remote control with the
help of IZYTRONIQ software.
IZYTRONIQ controls the test instrument in the remote mode, i.e.
sends control commands to the test instrument which then per-
forms the desired measurement. The terminal device with IZY-
TRONIQ is connected to the test instrument via USB.
The measurement data are transmitted from the test instrument
to IZYTRONIQ and thus do not appear at the test instrument’s
display.
Prerequisites
Remote control necessitates:
–
SECUTEST DB+ (Z853R or feature KB01)
(includes the following functions: “remote control via PC” and
“transfer user-defined test sequences from IZYTRONIQ to the
test instrument”)
–
IZYTRONIQ software
–
At least one user-defined IZY remote sequence
Note
Predefined test sequences cannot be executed via
remote control.
Procedure
➭
First of all, set up at least one user-defined IZY remote se-
quence in IZYTRONIQ. See IZYTRONIQ online help.
➭
Connect the test instrument to the PC with a USB cable (USB
slave interface, see section 5.4 “Controls and Connections”).
➭
Remote control operation is described in IZYTRONIQ online
help. Follow the instructions provided there.
Attention!
!
In the case of remote control or if a self-created IZY re-
mote sequence is used with the test instrument, the orig-
inator of the test sequences or the user/inspector
assumes responsibility for standards-compliant test
steps and execution of preliminary tests in the correct or-
der.
Attention!
!
Do not start any IZY remote sequences unless the test
instrument and the DUT are in plain view.
Do not connect line voltage to the test socket of your test
instrument before the surrounding area has been se-
cured.