106
Gossen Metrawatt GmbH
Test Management Errors
–
The current inspector is assigned
to the ETP role (electrically trained
person).
–
An ETP is not permitted to change
his own role or the role of another
inspector.
➭
Log in as a test planner in order to
change your own role or the role of
another inspector.
–
The current inspector is assigned
to the ETP role (electrically trained
person).
–
An ETP is not permitted to enter
rights for a new inspector.
➭
Log in as a test planner if you want
to enter a new inspector.
–
The current inspector is assigned
to the ETP role (electrically trained
person).
–
An ETP is not permitted to delete
an inspector’s rights.
➭
Log in as a test planner if you want
to delete an inspector.
Errors During Operation with Barcode or RFID Scanner
–
The scanned barcode is too long.
–
While writing an RFID tag an
attempt was made to write an ID to
the tag with vowel mutations such
as ä, ü or ö, or with special charac-
ters.
➭
Change vowel mutations such as ä
to ae.
➭
Avoid the use of special characters
in the ID.
Error Message
Possible Causes
Corrective Measures