Internal Database
A complete test structure with data regarding customer proper-
ties, buildings, floors, rooms and test objects can be created in
the test instrument.
Attention!
!
Sensitive Data – Mandatory Data Protection!
Customer data are confidential and must be protected.
Observe and comply with the respectively applicable na-
tional data protection regulations. Use the corresponding
functions provided by the test instrument such as pass-
word protection (section 11 “Inspector Management”),
as well as other appropriate measures.
This structure makes it possible to save the results of single mea-
surements or test sequences to test objects belonging to various
customers. Manual single measurements can be grouped
together into a so-called “manual sequence”.
Structures can be created directly at the test instrument or at a
PC (IZYTRONIQ software), and then transferred accordingly. They
can also be saved to and restored from a USB flash drive. See
section 19 “Transferring and Saving Test Structures and Measure-
ment Data (test instrument data base)”, for details.
Note
The scope of functions provided by the database struc-
ture and the transfer options depend on the instrument
variant and its features.
Database elements are grayed out on test instruments on
which the “extended database structure” (= property,
building, floor, room) from SECUTEST DB+ (Z853R or
feature KB01) or SECUTEST DB COMFORT (Z853S or
feature KD01) (medical electrical equipment) is not
enabled.
12.1
Test Structure Layout
The test structure is a tree structure with elements, i.e. the so-
called objects.
Objects
Objects can be identified with the following parameters (
boldface
parameters are mandatory entry fields):
•
Device
(
ID
, designation, location, test interval *, type, manufac-
turer, comment, serial number, protection class, cost center
*
,
department
*
)
•
ME equipment**
m
edical
e
lectrical equipment
(ID
, designation,
customer
, test interval*, type, manufacturer, comment,
serial number, protection class, number of type B applied parts **, num-
ber of type BF applied parts **, number of type CF ** applied parts, cost
center, department, UDI **, mains connection **)
•
Room *
(ID and
designation
)
•
Floor *
(ID and
designation
)
•
Building *
(ID,
designation
, street, ZIP code and city)
•
Property *
(ID and
designation
)
•
Customer
(
ID
,
designation
, street, ZIP code and city)
* Only with SECUTEST DB+ database extension (Z853R or feature KB01)
** Only with SECUTEST DB COMFORT database extension (Z853S or feature
KD01)
Up to 50,000 data records can be stored in the test instrument.
The following applies in this regard: 1 data record = 1 DUT or
location node or customer or individual measurement.
Note
Results of measurements/tests can be saved under
object types “Device” or “ME equipment”, which are also
referred to as “test objects” in the following. Measure-
ments can also be saved as a “manual sequence”.
Hierarchies
The following hierarchies must be complied with:
Room or Floor
must always be subordinate to a
Building
.
Devices or ME equipment
must always be subordinate to a
Customer
.
Note
Data Migration
If so-called “legacy data” have been imported into the
test instrument which do not comply with the hierarchy
(e.g. as a result of a firmware update or via the “Restore
database” function), customer objects are generated
automatically. The same applies to database objects
“Room” and “Floor”, which must always be subordinate
to a “Building”. In this case, Building objects are gener-
ated automatically if necessary.
Mandatory Entries
Mandatory entries are identified in red in the entry fields at the test
instrument as well as in the illustrations in this section.
Test Structure – Hierarchy of Object Levels in the SECUTEST ST BASE(10)
Figure 6
Database Structure
Customer
ID
Designation
Street
ZIP Code
City
Device
Database
ID
Designation
Measurement 1
Measurement 2
Measurement 3
Manual
Sequence
* Test interval: additionally available for test instruments with SECUTEST DB+ database extension (Z853R or feature KB01) or SECUTEST DB COMFORT
(Z853Sor feature KD01)
Customer MEM
1/4
2/4
3/4
4/4
ID
Test interval *
Comment
Cost center *
Designation
Type
Serial number
Department *
Location
Manufacturer
Protection class