![Gossen MetraWatt K852A Скачать руководство пользователя страница 42](http://html.mh-extra.com/html/gossen-metrawatt/k852a/k852a_operating-instructions-manual_2247272042.webp)
42
Gossen Metrawatt GmbH
4.8.2
Modifying, Adding, Inserting, Deleting Sequences
Exampel 1:
Based on the example 1 in section 4.8.1, step 1 modifies the time to 2 s. Add “CCH, 5 A, 1S” to the last step. Insert “CRL ,
1
Ω
,10 S” at step 3. Delete step 2. The sequence is chained to itself to realize the continuous execution.
Steps
Description
Display
Step 1
Press the
key +
key to enter into sequence operation menu.
►
No.: 0
Memo: Test Power
Step 2
Use the
▼
key to select Data:<New/Edit>.
►
Data:<New/Edit>
Count: 5
Step 3
Use the
►
key or knob to select Edit, and press the
Enter
key to confirm and enter step edit screen.
01. 1.00000s
CCL 1.0000A
Step 4
Use the
►
key to select time parameter, and the time set value flashes.
01. 1.00000s
CCL 1.0000A
Step 5
Use the Entry keys or the knob together with
◄
►
keys to input the time to 2 s, and press the
Enter
key to confirm.
01. 2.00000s
CCL 1.0000A
Step 6
Use the
▼
key to select the last new step (Step6).
06. 0.00000s
CRH 20.000
Ω
Step 7
Add a new step: CCH, 5 A, 1 S. The set value in the upper right corner of the display flashes. Use Entry keys or the knob together
with
◄
►
keys to input time to 1 s, and press the
Enter
key to confirm.
06. 1.00000s
CRH 20.000
Ω
Step 8
The basic mode showed in lower left corner of the display flashes. Use the knob or the
key to select operating mode to
CCH, and press the
Enter
key to confirm.
06. 1.00000s
CCH 0.000A
Step 9
The set value in the lower right corner of the display flashes. Use Entry keys or the knob together with the
◄
►
keys to input the
current value to 5 A. Press the
Enter
key to confirm and enter into the next operation.
06. 1.00000s
CCH 5.000A
07. 0.00000s
CCH 0.000A
Step 10
Use the
▲
key to select the step 3.
03. 1.00000s
CV 1.000V
Step 11
Press the
key +
key to insert a new step at step 3.
03. 0.00000s
CV 0.000V
Step 12
Insert a new step: CRL, 1
Ω
, 10 S at step 3. The setting operations for this new step are the same as described under items 7 to 9.
03. 10.00000s
CRL 1.0000
Ω
Step 13
Use the
▲
key to select the step 2.
02. 1.00000s
CCH 2.000A
Step 14
Press the
key +
key to delete the step 2.
02. 10.0000s
CRL 1.0000
Ω
Step 15
Press the
key to exit step edit screen and save the sequence data in EEPROM assigned by the sequence number.
►
Data:<New/Edit>
Count: 5
Step 16
Use the
▼
key to select Chain.
►
Chain: Off
Step 17
Use Entry keys or the knob to input 0 (it is chained to itself to realize continuous execution).
Press the
Enter
key to confirm and save the Chain value in EEPROM assigned by the sequence number.
►
Chain: 0
Step 18
Press the
key to exit sequence operation menu.
0.000V 0.000A
0.000A CCH
OFF