4.5.4
CP Mode
CP Mode includes constant power-current source mode (CPC) and constant power-voltage source mode (CPV).
Example 1:
Set the load to CPV and set the power value to 100 W.
Turn on the load, and check the present power value. The operating procedures are shown below:
For the above operating procedures, the corresponding SCPI commands should be:
Example 2:
Set the power value to 200 W in CPV mode.
Turn on the load, and there are two ways to set the power value.
Option 1:
Option 2:
For the above operating procedures, the corresponding SCPI commands should be:
Note
CPV /CPC will be shown respectively in the lower right corner of the display when the load is in CPV or CPC mode
Procedure
Operation Description
Display
1
Press the
key to enter into the mode selection and parameter setting menu.
MODE
:
CRL
RES
:
1.8000
Ω
2
Use the
▼
key to select CPV.
MODE
:
CPV
POWR
:
0.000W
3
Use Entry keys or the knob together with
◄
►
keys to input the power value to 100.
MODE
:
CPV
POWR
:
100.00W
4
Press the
Enter
key to confirm and exit the mode selection and parameter setting menu.
0.000V 0.000A
100.00W CPV
OFF
5
Press the
key to turn on the load.
0.000V 0.000A
100.00W CPV
ON
6
Check the present power value with
▲
and
▼
.
0.000V 0.000A
100.00W 0.000W
MODE
CPV
Sets the mode
POW
100
Sets the power value
INPUT
ON
Turns on the load
Procedure
Operation Description
Display
1
Press
key to enter into the mode selection and parameter setting menu.
MODE
:
CPV
POWR
:
100.00W
2
Use Entry keys or the knob together with
◄
►
keys to input the power value to 200.
MODE
:
CPV
POWR
:
200.00W
3
Press the
Enter
key to confirm and exit the mode selection and parameter setting menu.
0.000V 0.000A
200.00W CPV
OFF
4
Press the
key to turn on the load.
0.000V 0.000A
200.00W CPV
ON
Procedure
Operation Description
Display
1
Move the cursor to the hundredths with key
◄
.
0.000V 0.000A
100.00W CPV
OFF
2
Rotate the knob to set the hundredths to 2.
(Rotating the knob will change the set value immediately. The set value will become effective at input at once when the load is
turned on.)
0.000V 0.000A
200.00W CPV
OFF
3
Press the
key to turn on the load.
0.000V 0.000A
200.00W CPV
ON
MODE CPV
Sets the mode
POW 200
Sets the power value
INPUT ON
Turns on the load