34
Gossen Metrawatt GmbH
4.5.3
CR Mode
CR Mode includes constant resistance low range (CRL), constant resistance medium range (CRM), and constant resistance high range
(CRH).
Example 1:
Set the load to CRL and set the resistance value to 1.5
Ω
in CV mode.
Turn on the load, and check the present power value. The operating procedures are shown below:
For the above operating procedures, the corresponding SCPI commands should be:
Example 2
: Set the resistance value to 1.8
Ω
in CRL.
Turn on the load, and there are two ways to set the resistance value.
Option 1:
Option 2:
For the above operating procedures, the corresponding SCPI commands should be:
Note
CRH /CRM/CRL will be shown respectively in the lower right corner of the display when the load is in CRH status, CRM status
or CCL status.
Procedure
Operation Description
Display
1
Press
key to enter into the mode selection and parameter setting menu.
MODE
:
CV
VOLT
:
50.000V
2
Use
▼
key to select CRL.
MODE
:
CRL
RES
:
0.0200
Ω
3
Use Entry keys or the knob together with
◄
►
keys to input the resistance value to 1.5.
MODE
:
CRL
RES
:
1.5000
Ω
4
Press the
Enter
key to confirm and exit the mode selection and parameter setting menu.
0.000V 0.000A
1.5000
Ω
CRL
OFF
5
Press the
key to turn on the load.
0.000V 0.000A
1.5000
Ω
CRL
ON
6
Check the present power value with
▲
and
▼
.
0.000V 0.000A
1.5000
Ω
0.000W
MODE
CRL
Sets the mode
RES
1,5
Sets the resistance value
INPUT
ON
Turns on the load
Procedure
Operation Description
Display
1
Press the
key to enter into the mode selection and parameter setting menu.
MODE
:
CRL
RES
:
1.5000
Ω
2
Use Entry keys or the knob together with the
◄
►
keys to input the resistance value to 1.8.
MODE
:
CRL
RES
:
1.8000
Ω
3
Press the
Enter
key to confirm and exit the mode selection and parameter setting menu.
0.000V 0.000A
1.8000
Ω
CRL
OFF
4
Press the
key to turn on the load.
0.000V 0.000A
1.8000
Ω
CRL
ON
Procedure
Operation Description
Display
1
Move the cursor to the tenths with key
◄
.
0.000V 0.000A
1.5000
Ω
CRL
OFF
2
Rotate the knob to set the tenths to 8.
(Rotating the knob will change the set value immediately. The set value will become effective at input at once when the load is
turned on.)
0.000V 0.000A
1.8000
Ω
CRL
OFF
3
Press the
key to turn on the load.
0.000V 0.000A
1.8000
Ω
CRL
ON
MODE
CRL
Sets the mode
RES
1,8
Sets the resistance value
INPUT
ON
Turns on the load