Or via remote command to set:
Figure 2.6 shows the current waveform of the load: the load’s input current reaches the transient high level (10A) after 200 μs duration
of rising edge, and remains at 10A for 500 μs. Then after 200 μs duration of falling edge, the input current reaches the transient low
level (5A), and remains at 5A for 500 μs. Repeat it in cycles.
Figure 2.6
Continuous Transient Operation
2.4.2
Pulsed Transient Operation
The trigger function is required for pulsed transient operation. As long as no trigger occurs, the load remains at the transient low level.
After a trigger has been received, the load switches to high level and remains at this level for a programmed time span, subsequently,
the load returns to low level again. The associated parameters such as
– low level (LevelL),
– high level (LevelH),
– high level time (TimeH),
– time for rising edge (TimeR),
– time for falling edge (TimeF) and
– pulsed transient mode (pulse)
can be set through transient operation menu or via remote command
– (CURRent:LLEVel <NRf+>,
– CURRent:HLEVel <NRf+>,
– VOLTage:LLEVel <NRf+>,
– VOLTage:HLEVel <NRf+>,
– RESistance:LLEVel <NRf+>,
– RESistance:HLEVel <NRf+>,
– TRANsient:HTIMe <NRf+>,
– TRANsient:RTIMe <NRf+>,
– TRANsient:FTIMe <NRf+>,
– TRANsient:MODE PULSe).
Low level time (TimeL) has no influence on pulsed transient mode.
In order to get a pulse, an explicit trigger is required. The trigger can be an external trigger signal received via the TRIG input on the
front panel, pressing the
key and
key, via GPIB function
GET,
or via the common command
*TRG
or the subsystem com-
mand
TRIG
.
The trigger becomes effective only when the load remains at transient low level. Each trigger leads to one pulse. In the duration of rising
edge, transient high level, and falling edge, any trigger will be ignored.
Example: assume that the CCH range is active, and the input is in OFF status, then the transient parameters should be set as follows:
Press
key, to enter into transient operation.
Press
key, to open transient operation menu.
SCPI Command
Description
TRAN ON
Enables transient operation
CURR:LOW 5
Sets transient current low level to 5 A
CURR:HIGH 10
Sets transient current high level to 10 A
TRAN:LTIM 500 μs
Sets transient low level time to 500 μs
TRAN:HTIM 500 μs
Sets transient high level time to 500 μs
TRAN:RTIM 200 μs
Sets the time for transient rising edge to 200 μs
TRAN:FTIM 200 μs
Sets the time for transient falling edge to 200 μs
TRAN:MODE CONT
Selects continuous operation
INPUT ON
Turns on the power input