![Gossen MetraWatt K852A Скачать руководство пользователя страница 36](http://html.mh-extra.com/html/gossen-metrawatt/k852a/k852a_operating-instructions-manual_2247272036.webp)
36
Gossen Metrawatt GmbH
4.6
Short Circuit Operation
The operating procedures for short circuit operation are shown below:
Example 1:
The load was in CV mode previously and it is turned off. Set the load to short circuit in CCH status. Turn on the load to exe-
cute short circuit operation.
For the above operating procedures, the corresponding SCPI commands should be:
INPUT:SHORT
ON
Sets the load to short circuit operation
MODE
CCH
Sets the mode
INPUT
ON
Turns on the load
Example 2:
Based on the last example, exit the short circuit operation, and turn off the load.
For the above operating procedures, the corresponding SCPI commands should be:
INPUT:SHORT
OFF
Sets the load to exit the short circuit
INPUT
ON
Turns on the load
Procedure
Description
Display
1
Press the
key to enter into main menu, and select Short (short circuit) item with the
▼
key.
Short
:
On *Off
2
Use the
◄
key or knob to set On, and press the
Enter
key to confirm.
Short
:
On *Off
3
Press the
key to exit the main menu. The display shows “s” to indicate in the basic test mode.
0.000V
0.000A
0.000V
sCV OFF
4
Press the
key to enter into the mode selection and parameter setting menu; use
▲
and
▼
keys to select one basic
operating mode; press the
Enter
key to confirm and exit the mode selection and parameter setting menu.
0.000V 0.000A
44.000A sCCH
OFF
5
Press the
key to turn on the load.
0.000V 0.000A
44.000A sCCH
ON
Procedure
Description
Display
1
Press the
key to enter into main menu, and select Short (short circuit) item with the
▲
key.
Short
:
On *Off
2
Use the
◄
key or knob to set Off, and press the
Enter
key to confirm.
Short
:
On *Off
3
Press the
key to exit the main menu. On the display, “s” disappears in CCH status.
0.000V 0.000A
0.000A CCH
ON
4
Press the
key to turn off the load.
0.000V 0.000A
0.000A CCH
OFF