Functions and Features
2.1
Front Panel Operation („Local“) and Remote Control Operation
SPL Series electronic load can be controlled via the keypad and knobs in the front panel, or by remote controller via remote interface. If
it is needed to control the load via the front panel, the load has to stay in local control status. Local (front panel) control is in effect
immediately after power is applied. The REM indicator is turned on, and remote control goes into effect as soon as the load receives a
SYSTem:REMote command via RS232, USB or GPIB interface.
Under remote control status, all operations on front panel keypad and knobs become invalid (except
key and
key ). All
operations on electronic load are controlled by remote controller. The electronic load will return to local control and REM remote control
indicator is turned off after receiving the return command (SYSTem:LOCal).
Alternatively, you can return the electronic load to local control by pressing
key and
key.
Details of local operation are covered in Chapter 4 „Local Operation“ and fundamentals of remote programming are given in Chapter 5
„Remote Programming Operation“.
Complete SCPI programming details are given in the SCPI Programming Guide for SPL Loads.
2.2
Programmable Functions
•
CC (constant current) Mode: CCL, CCH
•
CV (constant voltage) Mode: CV
•
CR (constant resistance) Mode: CRL, CRM, CRH
•
CP (constant power) Mode: CPV, CPC
•
Transient Operation: Tran
•
List Operation: LIST
•
Battery Discharge Operation: BATTERY
•
Short Circuit Operation (Short)
2.3
Basic Test Functions
There are four basic test functions:
•
Constant current (CC),
•
Constant voltage (CV),
•
Constant resistance (CR),
•
Constant power (CP).
The operating mode and the associated parameters can be set via front panel or remote command. The load will remain in current
mode until the mode is changed. If the mode is changed when the load’s input is in ON status, the load will be turned off for around
5 milliseconds automatically.
The setting values for electronic load become effective immediately when the load is turned on. If the input set value exceeds the
allowed range, it will be automatically limited at maximum value or minimum value.