2.10
Measurement Function
The electronic load has measurement system with high resolution. The input current level and voltage level can be measured in real
time. The input power level and resistance level can be computed with the input voltage level and current level. Each measured value
can be checked through LCD display or via remote command (MEASure).
2.11
Saving and Recalling
The electronic load is provided with an EEPROM memory, which can save various parameters, such as modes, input status, current,
voltage, resistance, transient settings, limits, etc.
SPL Series electronic load can save 10 groups of parameters.
All parameters relevant to saving and recalling operation are listed in List 2-1.
The 10 groups of parameters stored in Location 0~9 can be saved and recalled by pressing
key +
key and
key +
key, or via the remote command (*SAV < NRl > and *RCL < NRl >).
The parameter saved in Location 0 will be recalled automatically every time the load is turned on.
List 2-1
* indicates it only can be programmed in the remote control.
2.12
Reading Remote Programming Errors
The Err indicator will be turned on when remote programming errors occur. The error codes are shown as followings:
Remote programming errors can be checked by pressing
key after pressing
key on the front panel. The remote command
(SYSTem:ERRor?) can read back the error codes and error messages when it is in remote control.
Function
Effect
Default
using the
example of
SPL 400-40
Input
Input status
Off
Mode
Operation mode
CCH
Current level
Immediate current level
0 A
Current rise rate
Current rise rate
4 A/μs
Current fall rate
Current fall rate
4 A/μs
Current Hlevel
Transient current high level
0 A
Current Llevel
Transient current low level
0 A
*Current protection level
Current limit
40 A
*Current protection delay
Current protection delay
60 s
*Current protection State
Enable/disable current protection
off
Voltage level
Immediate voltage level
80 V
CV current limit
Current limit in CV mode
40 A
Voltage Hlevel
Transient high voltage level
80 V
Voltage Llevel
Transient low voltage level
80 V
Resistance level
Immediate resistance level
2000
Ω
Resistance Hlevel
Transient high voltage level
2000
Ω
Resistance Llevel
Transient low voltage level
2000
Ω
Power level
Immediate power level
0 W
Transient operation
Transient test
off
Transient mode
Transient mode
continuous
Transient Htime
Time for transient high level
0 ms
Transient Ltime
Time for transient low level
0 ms
Transient Rtime
Time for transient rising edge
0.01 ms
Transient Ftime
Time for transient falling edge
0.01 ms
Trigger Function
Trigger function selection
Tran
Trigger source *
Trigger source
external
Battery Mode
Battery discharge operation
off
Battery mini voltage
Battery minimum termination voltage
0 V
Batterieentladestrom
Battery discharge current
0 A
Voltage on
Von point for the load
0 V
Voltage on Latch
Latch the Von point
Off
-lxx
Command errors
-2xx
Execution errors
-3xx
Device-specific errors
-4xx
Query errors