Gossen Metrawatt GmbH
19
Set the following parameters in transient operation menu:
Then press the
key to turn on the input.
Or via remote command to set:
Get the trigger by receiving an external trigger signal. Figure 2.7 shows the current waveform of the load before it is triggered and after
it has been triggered respectively: the electronic load starts its operation at the transient low level (5A) when the input is turned on. For
each trigger, the load current reaches the high level (10A) after 200 μs duration of rising edge, and remains at 10A for 500 μs. Then
after 200 μs duration of falling edge, the current returns to the transient low level (5A).
Figure 2.7
Pulsed transient operation
LevelL
5.000A
LevelH
10.000A
TimeH
0.50ms
TimeR
0.10ms
TimeF
0.10ms
Mode
Puls
SCPI Command
Description
TRIG:SOUR EXT
Selects the external trigger input
TRAN ON
Enables transient operation
TRAN:LLEV 5
Sets transient current low level to 5 A
TRAN:HLEV 10
Sets transient current high level to 10 A
TRAN:HTIM 500us
Sets transient high level time to 500 μs
TRAN:RTIM 100us
Sets the time for transient rising edge to 100 μs
TRAN:FTIM 200us
Sets the time for transient falling edge to 200 μs
TRAN:MODE PULS
Sets pulse trigger operating mode
INPUT ON
Turns on the input