38
Gossen Metrawatt GmbH
For the above operating procedures, the corresponding SCPI commands should be:
MODE
CV
Sets the mode
TRAN
ON
Enables the transient operation
VOLT:LOW
1
Sets a value to low level
VOLT:HIGH
5
Sets a value to high level
TRAN:LTIME
400 ms
Sets a value to low level time
TRAN:HTIME
200 ms
Sets a value to high level time
TRAN:RTIME
10 ms
Sets a value to time for rising edge
TRAN:FTIMR
20 ms
Sets a value to time for falling edge
TRAN:MODE
CONT
Selects continuous transient operation
INPUT
ON
Turns on the load
4.7.2
Pulsed Transient Operation
Example 1:
Assume that the load is in external triggering mode, set the load current value to periodically switch between 1 A and 5 A;
set the time for rising edge (TimeR) to10 ms; set high level time (TimeH) to 200 ms; set time for falling edge (TimeF) to 10 ms; the load
is in pulsed transient operation, and a trigger occurs at this time.
Procedure
Description
Display
1
Press the
key to enter into the mode selection and parameter setting menu; use the
▲ ▼
keys to select CCH
mode; press the
Enter
key to confirm and exit the mode selection and parameter setting menu.
0.000V
0.000A
0.000A CCH
OFF
2
Press the
Tran
key to enter into transient operation. The display shows “t” to indicate in CCH status.
0.000V
0.000A
0.500A tCCH
3
Press the
key to enter into the transient operation menu.
►
LevelL:
0.500A
LevelH:
1.000A
4
Use Entry keys or use the knob together with
◄
►
keys to set LevelL to 1, and press the
Enter
key to confirm.
LevelL: 1.000A
►
LevelH: 1.000A
5
Use Entry keys or use the knob together with
◄
►
keys to set LevelH to 5, and press the
Enter
key to confirm.
►
TimeL:
400.00ms
TimeH:
400.00ms
6
Press the
▼
key to select TimeH; use Entry keys or use the knob together with
◄
►
keys to set TimeH to 200 ms, and
press the
Enter
key to confirm.
►
TimeR:
10.00ms
TimeF:
20.00ms
7
Use Entry keys or use the knob together with
◄
►
keys to set TimeR to 10 ms, and press the
Enter
key to confirm.
TimeR:
10.00ms
►
TimeF:
20.00ms
8
Use Entry keys or use the knob together with
◄
►
keys to set TimeF to 10 ms, and press the
Enter
key to confirm.
►
Mode:
◄
Cont
►
9
Use the
◄
►
keys or knob to set Mode value to
pulse
, and press the
Enter
key to confirm.
►
Mode:
◄
Puls
►
10
Press the
key to exit the transient operation menu.
0.000V 0.000A
1.000A
tCCH OFF
11
Press the
key to turn on the load.
0.000V 0.000A
1.000A
tCCH ON
12
One trigger occurs when pressing
+
key or the signal at the trigger input terminal (TRIG IN) is low level.
0.000V 0.000A
5.000A
t
CCH ON