PART THREE - OPERATION
SECTION 6 - TEST/MAINTENANCE
Model F53 Flow Monitor/Totalizer (universal-mount 1/2 DIN)
Rev. 1-201
76
6.9 SELECT SIM
Measurement
You can simulate a measured value to make the relays and
analog outputs respond accordingly. First, select the type of
simulated value using this subsection. Then, set the desired
simulation value following the steps in subsection 6.10.
1.
With the
top-level menu screen dis-
played, use
Ø
Ø
key
to select the “SELECT SIM” line.
2.
Press
ENTER key
to display a screen like
. The number of sensors used deter-
mines the number of available choices. Use
Ø
Ø
and
×
×
keys
to view the choices:
•
FLOW A:
SIM value is Sensor A flow rate value.
•
FLOW B:
SIM value is Sensor B flow rate value.
•
FLOW C:
SIM value is Sensor C flow rate value.
•
FLOW D:
SIM value is Sensor D flow rate value.
•
VOLUME A:
SIM value is Sensor A volume value.
•
VOLUME B:
SIM value is Sensor B volume value.
•
VOLUME C:
SIM value is Sensor C volume value.
•
VOLUME D:
SIM value is Sensor D volume value.
•
CALCULATION*:
SIM value is calculated meas-
urement value (X-Y, X+Y, X/Y, or [X/Y] x 100).
*A simulated value can only represent the calculated measure-
ment when two sensors are used and the analyzer has been
correctly configured for CALCULATION.
3.
With the desired choice displayed, press
ENTER key
to enter this selection and return to the TEST/MAINT
top-level menu screen.