PART THREE - OPERATION
SECTION 6 - TEST/MAINTENANCE
Rev. 1-201
Model F53 Flow Monitor/Totalizer (universal-mount 1/2 DIN)
73
6.4 OVERFEED RESET
(relay timers)
6.5 OUTPUT (1 and 2)
Analog Test Signals
When a relay overfeed timer “times out,” as indicated by its
blinking annunciator, the timer must be manually reset using
TEST/MAINT menu screens. The relay annunciator stops
blinking after reset.
All overfeed timers are manually re-
set at once.
1.
With the
top-level menu screen displayed,
use
Ø
Ø
key
to select the “OVERFEED RESET” line.
2.
Press
ENTER key
to display “OVERFEED RESET”
screen. To reset all overfeed timers at once, press
ENTER key again (“OVERFEED RESET: DONE”
screen appears, acknowledging reset has occurred).
3.
To return to the TEST/MAINT top-level menu screen,
press
ESC key
or
ENTER key
.
The analyzer can provide analog output test signals of a
desired mA value to confirm operation of connected de-
vices.
These instructions only provide an Output 1 test
signal. Provide an Output 2 test signal in the same way
using its respective menu screens.
1.
With the
top-level menu screen dis-
played, use
Ø
Ø
key
to select the “OUTPUT 1” line.